Gill Batteries on Sale + Free Shipping on Select Models
Shop by Aircraft

Menu

PRO 360 - Kell-Strom KS5549 – PRO 360 Digital Protractor

Available for Order
Lead Time: Chat now or request quote
Specific Part Number

The PRO 360 is a specific part number of the series Kell-Strom KS5549. For more generalized information on this item series or for more options, click here.

CofC and 8130 available at no cost
Worldwide shipping available
Ask Our Aviation Specialists
Live Chat | +1.252.585.4553 | Email

Need a Quote?

Use this form to get a quote by email today.
First Name:*
Last Name:*
Email:*
Company Name: (optional)
Questions/Comments: (optional)
Alternative Model Number for PRO-360 Digital Protractor
Alternate Models
09429606Alternative Part Number
PRO 360Alternative Model Number for PRO-360 Digital Protractor
General Information
Part #PRO 360
ManufacturerKell-Strom
NSN4920014206291
Shipping Information
Product Weight1.5 lbs.
Product Dimensions8.3 x 5.3 x 2.3 in.
Schedule B9017208060
ECCNEAR99

Customers also viewed

Tronair KHC-4046
Kit, Qd Adaptor
Kell-Strom KS5518E
European Version – 230/220V
Kell-Strom KS5518
Electrosonic Cleaner – 117V/110V
Kell-Strom KS5549-WC
PRO 360 Digital Protractor (With Calibration)
Kell-Strom KS6005-WC
– 0.01° Accuracy – RS-232 Serial Port

Customers who bought this item also purchased

Tronair 02-7813C0100
with Spacer Blocks, 4.5" to 15", CE-Marked
Pacific Scientific T5-2002-401-00
Dual-Calibration: 10–70 CWT British & 1/16"–1/4" Inch; 10–150 lb Range; Cal Cards T594-101 & T610-401B
Tronair 02-7835C0100
8.4" to 27.9", CE-Marked – Select Bombardier/Dassault Aircraft
Tronair 14-6806-6011
Aircraft Tire Pressure Gauge, 0 to 300 psi
Pacific Scientific T60-1001-C8-1A
10–200 lbs; 1/16"–1/4" Cables; No Risers Required; Built-In Cable Gauge; USAF C8 Equivalent

AIRCRAFT COMPATIBILITY

This item is for use on all aircraft.

PRODUCT REVIEWS

This item does not have any reviews yet.

QUESTIONS & ANSWERS

Question:
How often should I calibrate the PRO 360 digital protractor?
Answer:

To maintain accurate measurements, the PRO 360 should be tested for accuracy daily using the “end-to-end” method. If any of the four readings differ by more than 0.1°, recalibration using the Superset® procedure is required. This is especially important after drops or when operating in environments with temperature shifts greater than 9°F (5°C) from the last calibration setting.

Asked June 27, 2025
Question:
What is the Superset® calibration method?
Answer:

Superset® is a built-in, eight-step electronic recalibration process that realigns the PRO 360 across its entire 360° range. It records four horizontal and four vertical positions, ensuring precision without special tools. The process takes only a few minutes and can be done onsite by following the step-by-step instructions in the owner’s manual.

Asked June 27, 2025
Question:
What kind of battery does the PRO 360 use, and how long does it last?
Answer:

The PRO 360 runs on a 9-volt alkaline battery, which provides up to 500 hours of typical use. A lithium 9V battery can be used for extended life. The unit features an automatic shutoff after five minutes of inactivity and a low battery warning that flashes “LO bAt” when it’s time to replace it.

Asked June 27, 2025
Question:
Does the PRO 360 store measurements?
Answer:

While it doesn’t have permanent memory storage, the PRO 360 includes a HOLD feature that temporarily “freezes” the current reading. This is useful when measuring in hard-to-read positions or when recording values. Pressing the HOLD button again will release the display for continued measurements.

Asked June 27, 2025
Question:
Can I set a custom zero point on the PRO 360?
Answer:

Yes, the ALT ZERO function allows you to designate any angle as your reference 0.0° point. This is helpful when measuring relative angles from surfaces that aren’t perfectly level. Simply place the device on the reference surface, wait five seconds, and press ALT ZERO to activate the alternate mode

Asked June 27, 2025
Question:
What kind of maintenance does the PRO 360 require?
Answer:

Routine maintenance is minimal. Clean the unit with mild liquid soap and a damp cloth – never immerse it in water or use solvents. Perform the daily accuracy test before use, and recalibrate as needed. Proper storage and care will extend the unit’s service life.

Asked June 27, 2025
Question:
What warranty is offered on the PRO 360?
Answer:

The PRO 360 digital protractor includes a one-year limited warranty covering defects in materials and workmanship. It does not cover damage from misuse, tampering, water exposure, or improper maintenance. Warranty service requires proof of purchase and a brief explanation of the issue.